Design, performance, and calibration of the CMS hadron-outer calorimeter


Creative Commons License

Abdullin S., Abramov V., Acharya B., Adam N., Adams M., Akchurin N., ...Daha Fazla

European Physical Journal C, cilt.57, sa.3, ss.653-663, 2008 (SCI-Expanded) identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 57 Sayı: 3
  • Basım Tarihi: 2008
  • Doi Numarası: 10.1140/epjc/s10052-008-0756-6
  • Dergi Adı: European Physical Journal C
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.653-663
  • Yıldız Teknik Üniversitesi Adresli: Evet

Özet

The Outer Hadron Calorimeter (HCAL HO) of the CMS detector is designed to measure the energy that is not contained by the barrel (HCAL HB) and electromagnetic (ECAL EB) calorimeters. Due to space limitation the barrel calorimeters do not contain completely the hadronic shower and an outer calorimeter (HO) was designed, constructed and inserted in the muon system of CMS to measure the energy leakage. Testing and calibration of the HO was carried out in a 300 GeV/c test beam that improved the linearity and resolution. HO will provide a net improvement in missing E T measurements at LHC energies. Information from HO will also be used for the muon trigger in CMS. © 2008 Springer-Verlag / Società Italiana di Fisica.